Maintenance measures for dental equipment | dental operating chair



Because dental machines are necessary equipment for dentists to treat patients, the usage rate is high. It is also a device that integrates water, gas, and electricity. Daily maintenance is directly related to the life of the equipment and the intact rate of the equipment. Therefore, it is particularly important to stick to the daily maintenance of the tooth machine.
    1. The first is the maintenance of the mobile phone, which is a key component of the dental comprehensive treatment chair. During use, in order to avoid cross-infection between patients, the mobile phone must be sterilized by high temperature and high pressure. A vacuum sterilizer is recommended. In order to prolong the service life of the mobile phone, special attention should be paid to daily maintenance. Before the first use, use the injection needle to drop 1 to 2 drops of high-speed lubricating oil on the shaft (recommended: ORAL-E tooth cavity company's mobile phone cleaning lubricating oil). Under normal circumstances, the front of the car should be cleaned and lubricated daily Clean the micro-bearing once every 2 weeks of work. The normal working pressure of 0.2~0.25Mpa should be maintained; when there is no water, the mobile phone can not be idling, otherwise the bearing will be damaged. When the bur is blunt, replace it with a new bur in time, otherwise it will affect the life of the bearing. It is recommended to use the mobile phone oiling cleaning machine for cleaning after using the mobile phone, which can clean 3 mobile phones at the same time. The unique cleaning mode saves time and oil. The unique oil collection drawer saves a lot of time for the doctors in the dental hospital to save a lot of time during the busy work. It is recommended to use Matching oil injection machine cleaning oil (set).
    2. The maintenance of the tooth machine is mainly the maintenance of the air circuit and the water circuit.
    a. It is recommended that the gas, water and electricity should be turned off every day after get off work. It is particularly worth mentioning that after turning off the gas, a three-purpose spray gun should be used to exhaust the air in the lumen to prevent the pipeline from being in a high-pressure state for a long time and easy to aging and rupture.
    b. Clean the water filter regularly according to the local water quality, and regularly drain the water in the air filter pressure reducing valve inside the tooth machine.
    c. Clean the powerful suction device and saliva suction device inside the dental machine in time, and often take clean water to clean the pipeline of the dental machine.
    d. The operating lamp and film viewing lamp of the dental machine cannot be switched on and off frequently, which will shorten the life of the bulb.
    e. Correctly adjust the water pressure and air pressure of the tooth machine, and must be within the specified range, so as not to cause the filter and pipeline to burst.
    f. The tooth machine is still a mechanical mechatronics equipment, it is recommended to lubricate the equipment for about half a year
    3. In the maintenance of the patient seat, each transmission component should be checked regularly, and the mechanical transmission position should be regularly lubricated. Electric hydraulic transmission should regularly check the oil quantity and oil quality in the oil tank, and timely add and replace when the oil quantity is insufficient or the oil quality is not good. The belt drive should pay attention to adjusting the tightness of the belt.